DBTEC-2A
Thermoelectric Controller |
|
|
|
Specifications
& Dimensions
|
TE Controller Output . |
|
Control Loop: |
PI |
|
Current range: |
+
2 Amps |
Compliance voltage (Iout = 1 Amp):
Compliance voltage (Iout
= 2 Amp): |
|Vs - 2.0| Volts
|Vs - 2.6| Volts |
|
Temperature Sensors . |
|
Types: |
NTC (2 wire), RTE's, AD590, LM335 |
|
Input . |
|
Offset Voltage,
Initial: |
1 mV |
|
Bias Current: |
20 nA |
|
Offset Current: |
2 nA |
|
Common Mode Range: |
VDD - 1.5 Volt |
|
Input Impedence: |
500 kW |
|
Power Supply . |
|
Voltage, VDD: |
+5 VDC to +12 VDC |
|
Current, VDD supply: |
55 mA |
|
Voltage, VS: |
+5 VDC up to +12 VDC |
|
Current, VS supply: |
2.0 Amps max. |
|
Thermal . |
|
Heatspreader temperature
rise: |
+50 deg. C/W |
|
Heatspreader with
heatsink temperature rise: |
+21.5 deg. C/W |
Heatspreader with
heatsink and fan
temperature rise: |
+3.4 deg. C/W |
|
|
Mounting Methods
The DBTEC-2A can be mounted
in two ways. The first uses a heat sink mounted on the TEC board
and uses air cooling from instrument fan. In situations where
there is no instrument fan, a small fan can be mounted directly
to the TEC board. The second method mounts the TEC onto instrument
chassis or large heat sink in instrument. Air flow is highly
recommended to keep the heat sink cool. |
|

Method 1
Heat sink mounted on board. Use instrument fan for air cooling.
Method 2
Board is mounted onto instrument chassis or large heat sink.
|
|
Board Set Up
 
|
|
Dimensions

|
|
Ordering Information
|
Model # |
Description |
|
DBTEC-2A |
+
2 Amp Thermoelectric Controller |
|
DBKIT |
Contains heat sink, thermal pad,
screw driver, stand-off |
|
DBHS |
Heat sink and thermal pad for controller |
|
DBFAN-5V |
5 Volt fan for controller |
|
DBFAN-12V |
12 Volt fan for controller |
|
|